Chromatography Fundamentals
The core principles behind every separation: how columns work, what drives selectivity, and how to think about resolution.
What is chromatography?
Chromatography separates molecules by partitioning them between a stationary phase (the column matrix) and a mobile phase (the buffer flowing through it). Molecules that interact more strongly with the stationary phase travel more slowly — different affinities produce different elution times.
The van Deemter equation
Column efficiency is described by the height equivalent to a theoretical plate (HETP):
Where u is linear flow rate, A accounts for eddy diffusion,B for longitudinal diffusion, and C for mass transfer resistance. Velocity conventions and coefficients must agree: superficial velocity is Q/Acol, whereas interstitial velocity is Q/(εiAcol), with Acol the column cross-sectional area and εi the interparticle void fraction. The simulator below uses interstitial velocity. This is a simplified efficiency model; its useful operating range and the observed optimum depend on the column, mobile phase, analyte, and instrument. Evaluate performance against the selected column’s instructions and a defined test method.
Resolution
For two resolved features, report retention coordinates and base widths consistently:
- tR — retention time; calculate both peaks in the same time coordinate (or consistently transform both to volume).
- wb — base width in that same coordinate, measured under a stated convention.
- Rs — a description of the observed pair of peaks, not proof of identity or a pure collected fraction.
The familiar Rs ≈ 1.5 “baseline separation” rule is an idealized heuristic for suitably shaped, similarly responding peaks—not a universal pooling or purity guarantee. The plate/selectivity/retention relationship shown in many textbooks is an isocratic model with stated assumptions; do not apply it as a general gradient-elution equation. In practice, compare methods using measured resolution and the decision’s relevant orthogonal assay.
Plate count
A practical way to measure column health. For a symmetric peak eluting at retention volume VRwith peak width at half-maximum w½:
Record the test solute, flow, detector, peak-width convention, and whether the result is reported as plates or plates per metre. Vendor specifications are method- and column-specific; a change in measured efficiency warrants investigation but does not by itself identify fouling or a void.
Plate height, plate number and resolution (idealized model)
Plate height follows the reduced van Deemter form h = a + b/ν + c·ν with h = H/dp and ν = u·dp/Dm (u is interstitial velocity). The optimum is νopt = √(b/c), hmin = a + 2√(bc). The bed then gives N = L/H, and the lower plot shows two equal-amount Gaussian peaks eluting isocratically at VR/V0 = 1 + k with the same N, and Rs computed from their base widths (4σ).
- H
- 6.55 µm
- N = L/H
- 22,900
- Reduced velocity ν
- 5
- u / u_opt
- 0.833
- H_min
- 6.5 µm
- Rs (base width)
- 2.44
c between 0.02 and 0.05 is the range quoted for well-packed columns; larger values illustrate slower mass transfer. For large proteins in porous media, ν is high at usual flows and the C term dominates.
Idealized: one plate height for both species, Gaussian peaks, isocratic retention, no extra-column broadening; a and b fixed at 0.7 and 1.8, Dm values are orders of magnitude only. It does not describe gradient elution or predict a real column. Reduced van Deemter form, optimum, typical coefficients and N = L/H from Pirok & Schoenmakers, Analytical Separation Science course, lesson 03 (ass-ets.org, read 2026-09-25); base-width Rs as on this page.
Peak asymmetry and tailing
Real peaks are not perfect Gaussians. An asymmetry calculation, such as As = b/a at a stated percentage of peak height, is only comparable when the convention and test conditions are recorded. Tailing or fronting can arise from several causes—including sample load, solvent mismatch, extra-column effects, kinetics, or column/sample interactions—and needs troubleshooting evidence rather than a fixed acceptance boundary or a single-cause diagnosis.
Gradient elution vs. isocratic
In isocratic elution, mobile phase composition is constant — common in SEC. In gradient elution (IEX, HIC, RPC), a changing mobile phase progressively elutes bound molecules: increasing salt in IEX, decreasing salt in HIC, increasing organic modifier in RPC.
Gradient slope can change selectivity, retention, peak compression, run time, and dilution. Compare candidate gradients experimentally under a defined objective; shallower or steeper is not universally better.
Sources and scope
IUPAC Gold Book: peak resolution (authoritative definition, accessed 2026-09-25) supports the displayed base-width equation. Cytiva™ Superdex™ 200 Increase instructions for use (official, accessed 2026-09-25) illustrates that a specific column’s efficiency test is reported as plates per metre under stated conditions. Neither source supports a universal resolution threshold, gradient equation, plate-count specification, or peak-asymmetry acceptance band. The simulator’s reduced van Deemter form, optimum, typical coefficients and N = L/H follow Pirok and Schoenmakers, Analytical Separation Science, lesson 03 (course notes, accessed 2026-09-25); its numbers are illustrative.