Troubleshooting Guide
Diagnose the most common chromatography problems — peak tailing, poor resolution, ghost peaks, high pressure, and baseline drift.
Systematic troubleshooting starts with reproducing the problem, then isolating the variable. Change one thing at a time — run a blank (no sample) first to distinguish system artefacts from sample-related issues.
Peak tailing
Likely causes
- Column packed too loosely or bed damaged
- Non-specific ionic or hydrophobic interactions with the medium (for example in SEC)
- Target not binding properly under the start conditions (IEX/HIC)
- Extra-column volume in tubing and connections
What to try
- Measure asymmetry (As = b/a at 10% of peak height) and efficiency, and compare with earlier tests of the same column on the same system
- In SEC, adjust buffer composition, for example ionic strength, pH or an additive, within the column limits
- Check start-buffer pH and salt, and whether the load is within capacity
- Minimise tubing length and connections between injector, column and detector
Poor resolution between adjacent peaks
Likely causes
- Gradient too steep for the separation (IEX/HIC)
- Flow rate above the range that gives adequate resolution
- Loss of column performance (packing, fouling)
- Sample volume or load too high
- Selectivity mismatch (wrong resin or conditions for the separation)
What to try
- Try a shallower gradient over the elution window
- Test a lower flow rate within the column specification and compare resolution
- Clean the column, then re-test efficiency; efficiency depends on the system, so compare with the column baseline rather than a fixed fraction of the specification
- Reduce sample volume or load and compare
- Scout different pH, salt or an alternative resin chemistry
Ghost peaks or extra peaks
Likely causes
- Buffer impurities concentrated on the column and released during elution
- Material retained from a previous run
- Sample and running buffer mismatch causing refractive-index disturbances in the UV trace (often negative peaks near the solvent front)
- Air bubble passing through the UV flow cell (spikes)
What to try
- Run a blank gradient with no sample to see which peaks come from the system or buffers
- Clean the column as described in its instructions
- Exchange or dilute the sample into start or running buffer
- Use degassed buffers and check the pump and inlets for air
High system pressure
Likely causes
- Column or column filter fouled or blocked
- Particulates in sample or buffer
- Viscous sample or buffer, or low temperature
- Flow rate too high for the column and conditions
- Blockage in system tubing, valves or in-line filters
What to try
- Clean the column with the procedure in its instructions; use reverse flow only if the column instructions allow it
- Filter or centrifuge samples and filter buffers, choosing filter pore size to suit the medium; check recovery if the target may adsorb to the filter
- Dilute viscous samples and reduce flow for viscous buffers or cold operation
- Check the column instructions for pressure and flow limits, and set system alarm limits accordingly
- Run the system without the column to locate where the pressure arises
Baseline drift
Likely causes
- Buffers A and B absorb differently at the monitored wavelength
- Temperature changes in the lab, buffers or column
- Contamination or deposits in the UV flow cell
- Air in the flow cell
What to try
- Balance UV absorbance of buffers A and B, or monitor at a different wavelength
- Let instrument, column and buffers reach the same temperature before the run
- Clean the flow cell following the system manual
- Degas buffers and purge air as described in the system manual
Sources and scope
Causes and remedies follow the troubleshooting tables in Cytiva™’s Size Exclusion Chromatography, Ion Exchange Chromatography and Hydrophobic Interaction and Reversed Phase Chromatography handbooks (official, accessed 2026-09-25). They are diagnostic prompts, not fixed recipes: take flow, pressure, cleaning and buffer limits from the instructions for the exact column, resin and system.