Troubleshooting Guide

Diagnose the most common chromatography problems — peak tailing, poor resolution, ghost peaks, high pressure, and baseline drift.

Systematic troubleshooting starts with reproducing the problem, then isolating the variable. Change one thing at a time — run a blank (no sample) first to distinguish system artefacts from sample-related issues.

Peak tailing

Likely causes

  • Column packed too loosely or bed damaged
  • Non-specific ionic or hydrophobic interactions with the medium (for example in SEC)
  • Target not binding properly under the start conditions (IEX/HIC)
  • Extra-column volume in tubing and connections

What to try

  • Measure asymmetry (As = b/a at 10% of peak height) and efficiency, and compare with earlier tests of the same column on the same system
  • In SEC, adjust buffer composition, for example ionic strength, pH or an additive, within the column limits
  • Check start-buffer pH and salt, and whether the load is within capacity
  • Minimise tubing length and connections between injector, column and detector

Poor resolution between adjacent peaks

Likely causes

  • Gradient too steep for the separation (IEX/HIC)
  • Flow rate above the range that gives adequate resolution
  • Loss of column performance (packing, fouling)
  • Sample volume or load too high
  • Selectivity mismatch (wrong resin or conditions for the separation)

What to try

  • Try a shallower gradient over the elution window
  • Test a lower flow rate within the column specification and compare resolution
  • Clean the column, then re-test efficiency; efficiency depends on the system, so compare with the column baseline rather than a fixed fraction of the specification
  • Reduce sample volume or load and compare
  • Scout different pH, salt or an alternative resin chemistry

Ghost peaks or extra peaks

Likely causes

  • Buffer impurities concentrated on the column and released during elution
  • Material retained from a previous run
  • Sample and running buffer mismatch causing refractive-index disturbances in the UV trace (often negative peaks near the solvent front)
  • Air bubble passing through the UV flow cell (spikes)

What to try

  • Run a blank gradient with no sample to see which peaks come from the system or buffers
  • Clean the column as described in its instructions
  • Exchange or dilute the sample into start or running buffer
  • Use degassed buffers and check the pump and inlets for air

High system pressure

Likely causes

  • Column or column filter fouled or blocked
  • Particulates in sample or buffer
  • Viscous sample or buffer, or low temperature
  • Flow rate too high for the column and conditions
  • Blockage in system tubing, valves or in-line filters

What to try

  • Clean the column with the procedure in its instructions; use reverse flow only if the column instructions allow it
  • Filter or centrifuge samples and filter buffers, choosing filter pore size to suit the medium; check recovery if the target may adsorb to the filter
  • Dilute viscous samples and reduce flow for viscous buffers or cold operation
  • Check the column instructions for pressure and flow limits, and set system alarm limits accordingly
  • Run the system without the column to locate where the pressure arises

Baseline drift

Likely causes

  • Buffers A and B absorb differently at the monitored wavelength
  • Temperature changes in the lab, buffers or column
  • Contamination or deposits in the UV flow cell
  • Air in the flow cell

What to try

  • Balance UV absorbance of buffers A and B, or monitor at a different wavelength
  • Let instrument, column and buffers reach the same temperature before the run
  • Clean the flow cell following the system manual
  • Degas buffers and purge air as described in the system manual

Sources and scope

Causes and remedies follow the troubleshooting tables in Cytiva™’s Size Exclusion Chromatography, Ion Exchange Chromatography and Hydrophobic Interaction and Reversed Phase Chromatography handbooks (official, accessed 2026-09-25). They are diagnostic prompts, not fixed recipes: take flow, pressure, cleaning and buffer limits from the instructions for the exact column, resin and system.

Try it on a trace

Open a two-column chromatogram in the workbench, or load the synthetic example. No account needed.