Flow rate, linear velocity and residence time

Convert between volumetric flow (mL/min), linear velocity (cm/h), residence time (min) and column volumes per hour for a packed column.

A pump is set in mL/min. Linear velocity in cm/h is that flow divided by the column cross-section, so the same value means the same flow per unit area on any diameter. Dynamic binding capacity can depend on residence time. These are the same flow seen three ways, linked by the column geometry.

Convert flow, linear velocity and residence time

Enter the column geometry, then type into any one of the four flow fields. The other three follow. The field you typed into stays fixed when you change the geometry.

Flow quantities

Fixed: Volumetric flow Q. Values are rounded to 4 significant digits.

Column volume CV = π(d/2)²·h
20.11 mL
Cross-sectional area A
2.011 cm²
Flow Q
2 mL/min
Linear velocity u
59.68 cm/h
Residence time τ
10.05 min
Column volumes per hour
5.968 CV/h
From the resin or column instructions, for your liquid and temperature.

u = Q / A × 60 (cm/h, with Q in mL/min and A in cm²). τ = h / u × 60 (min) = CV / Q. CV/h = 60 / τ.

τ is empty-bed residence time: packed bed volume divided by flow. Pressure limits depend on the resin, the column hardware and the liquid viscosity, which changes with temperature and additives. This tool does not check pressure. Nothing you enter is saved or sent.

The formulas

For a column of inner diameter d and bed height h, both in cm:

  • Cross-sectional area: A = π (d/2)², in cm².
  • Column volume: CV = A · h = π (d/2)² · h, in mL (1 cm³ = 1 mL).
  • Linear velocity: u = Q / A × 60, in cm/h, with Q in mL/min. The factor 60 converts minutes to hours.
  • Residence time: τ = h / u × 60 = CV / Q, in min.
  • Column volumes per hour: CV/h = 60 / τ.

Linear velocity here is superficial velocity: flow divided by the empty column cross-section. Liquid moves faster between the beads, because only part of the bed is open to flow. Residence time is likewise the empty-bed value, packed bed volume divided by flow. Some papers use the void volume instead, which gives a shorter time. State which one you use.

Worked example

A column with 1.6 cm inner diameter and a 10 cm bed: A = π × 0.8² = 2.01 cm², and CV = 20.1 mL. At Q = 2 mL/min, u = 2 / 2.01 × 60 = 59.7 cm/h. τ = 20.1 / 2 = 10.05 min, and CV/h = 60 / 10.05 = 5.97. These are the calculator defaults.

Which number to hold constant

  • Linear velocity compares runs on columns of different diameter but the same bed height, because it is divided by the cross-section. For the same reason, a flow limit given in cm/h applies to any diameter, while one given in mL/min applies only to the column it was written for. Check whether your resin and column instructions state the limit in cm/h, in mL/min or as a pressure.
  • Residence time is the contact time between liquid and bed. Dynamic binding capacity can depend on it, because uptake into porous beads takes time; how much is specific to the resin and feed. Measure it as in measuring dynamic binding capacity.
  • CV/h is residence time in another form. It is convenient when step lengths are written in CV.

To move a method to a larger column, see scale up at constant residence time and bed height.

Limits this tool does not check

Pressure drop over a packed bed grows with linear velocity, bed height and liquid viscosity. It also depends on the resin particle size, how the bed was packed and the column hardware. Cold buffers and additives such as glycerol raise viscosity and pressure. Use the resin and column instructions for velocity and pressure limits under your conditions. If the limit is in cm/h, enter it above to get a warning when the computed velocity exceeds it; convert a limit in mL/min with the converter first.

The example geometries are generic inner diameter × bed height pairs for common lab formats. Prepacked columns list their exact bed volume and dimensions on the data sheet; use those values.

Sources and scope

The formulas are geometry; the page text is original. Results are not saved and are not sent anywhere.

Try it on a trace